SONG spectrograph assembly
- Prepare optical table with acceptable support and sufficient access on all sides (while attempting to minimize total footprint
- Mark optical table with locations of mounts based on lab measurements
- set up upper table on main optical bench in correct location. Install:
- fiber input,
- collimating assembly. Check with auto-collimator.
- laser input via removable beam splitter,
- camera output via removable beam splitter, Check with auto-collimator?
- output focusing lens.
- Check alignment with laser, ensure camera functionality.Install slit to ensure focal plane is at desired height above optical table
- Install FM1 at desired height. Position to get desired deflection angle
- Install collimating telescope behind table. Adjust diagonal mirrors to ensure proper alignment
- Install grating mirror and align with collimating telescope
- Install FM2 and align with collimating telescope (do now to ensure consistency with grating mirror)
- Install FM3 and align with collimating telescope (do now to ensure consistency with grating mirror)
- Install OAP1. Adjust initial height and tilt using autocollimator off mirror on back of mount. Adjust lateral so laser falls at appropriate location on the OAP (52mm from edge?).
- Align OAP1 using 10micron fiber input to minimize aberration
- Rotate spectrograph mirror to get 15mm separation between input and output laser. Adjust further to ensure that reflected beam misses FM1.
- Install OAP2. Adjust initial height and tilt using autocollimator off mirror on back of mount. Adjust lateral so laser falls at appropriate location on the OAP.
- Rotate FM3 so that return beam goes back into camera.
- Align OAP2 using 10 micron fiber input to minimize aberration
- Install prism. Rotate FM3 to ensure correct beam location
- Install camera. Adjust to ensure correct beam location: no light falling outside any optic
- Image cross-disperser fiber. Ensure expected width. Adjust detector to make vertical spectrum. Rough prism adjustment to get laser at correct location on detector.
- Install grating.
- Adjust grating tilt and prism to ensure good spectral coverage: grating so that peak of blaze function near center of detector, and prism to ensure good wavelength coverage.