| ====October 31 2026==== | ====October 31 2026==== | ||
| - | - QHY camera apparently recovered! although lots of ups and downs along the way, Hypothesis: | + | - QHY camera apparently recovered! although lots of ups and downs along the way, with it seeming to fail, but eventually got it working (not that software |
| - | - investigated watchdog system and recognized that perhaps relay trigger needs to be on for more than one second to trigger watchdog reset.Changed to 5 seconds and testing... | + | - investigated watchdog system and recognized that perhaps relay trigger needs to be on for more than one second to trigger watchdog reset.Changed to 5 seconds and testing...seems to be holding. |
| - cal fiber shutter not working due to apparent relay failure. Replaced LCUS relay with Noyito relay | - cal fiber shutter not working due to apparent relay failure. Replaced LCUS relay with Noyito relay | ||
| - | - swapped purple wire thermocoupld | + | - swapped |
| + | |||
| + | - checked lower dome, wouldn' | ||
| + | |||
| + | ====October 20 2026==== | ||
| + | |||
| + | Catastrophic (?) failure of chiller. CCD held temperature about 10 minutes as thermocouples rose, then CCD power hit 100% and CCD temperature rose over 10 minutes, then perhaps power set to zero, then no response from CCD. Couldn' | ||
| ====August 2026==== | ====August 2026==== | ||