on_sky_testing
Differences
This shows you the differences between two versions of the page.
| Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
| on_sky_testing [2026/06/02 14:59] – holtz | on_sky_testing [2026/06/03 13:25] (current) – holtz | ||
|---|---|---|---|
| Line 15: | Line 15: | ||
| - inspected Shelyak calibration source due to no LED light and determined that indeed, no LEDs appear to be coming on inside the unit. Emailed Shelyak. | - inspected Shelyak calibration source due to no LED light and determined that indeed, no LEDs appear to be coming on inside the unit. Emailed Shelyak. | ||
| + | |||
| + | - took pixel flats with liquid cooled QHY600. This was achieved using a LED penlight wrapped in some bubble wrap and inserted into the upper light feed without moving anything in the spectrograph. Took 101 .0005s exposures | ||
| + | |||
| + | - checked lower dome, which wouldn' | ||
| + | |||
| + | - at end of run, inspected calibration frames and see apparent drop in flux from previous week, especially in flats, but also in ThAr, with drop in both cal channel and direct channel. Lots of investigation followed, determined that fluxes can change with reseating of fiber in cal source, although still couldn' | ||