| + | ====June 2026==== | ||
| + | |||
| + | Various work: | ||
| + | |||
| + | - the telescope started to oscillate in the past several weeks, evident in the Planewave tracking plots. We ran a quick tuning, but no change. We ran a thorough tuning (couple hours), also no change, but upon rebooting pwi1m computer, problem seems to have gone away. Consulting Planewave (Josh Persson), this is not expected, so unclear what was/is going on! | ||
| + | |||
| + | - we installed a new stage for the iodine cell; previous stage was damaged and shipped to Thorlabs for repair. We used the bright LED source through a 10-micron fiber in the cal channel to project the beam through the system. We attached a card to front of iodine cell with a cross in the middle, and attempted | ||
| + | to position the cell (in height) to center the beam on the spot; subsequently, | ||
| + | |||
| + | - using 10 micron fiber with calibration source, investigated cal channel image quality, and image quality through iodine cell. Found that cal channel image quality was very poor. Attempted adjustment of both collimator and camera lenses and was unable to make any significant improvements. NOTE; we don't really care about image quality in cal channel because it is a diffuse source for calibration observations, | ||
| + | |||
| + | - redetermined calibration spot and hole positions. Focussed guide camera. | ||
| + | |||
| + | - connected the QHY600 USB directly to the computer, rather than running through the Wanderer, in an attempt to cure the disconnects we have been getting | ||
| + | |||
| + | - inspected Shelyak calibration source due to no LED light and determined that indeed, no LEDs appear to be coming on inside the unit. Emailed Shelyak. | ||
| + | |||
| + | - took pixel flats with liquid cooled QHY600. This was achieved using a LED penlight wrapped in some bubble wrap and inserted into the upper light feed without moving anything in the spectrograph. Took 101 .0005s exposures | ||
| + | |||
| + | - checked lower dome, which wouldn' | ||
| + | |||
| + | - at end of run, inspected calibration frames and see apparent drop in flux from previous week, especially in flats, but also in ThAr, with drop in both cal channel and direct channel. Lots of investigation followed, determined that fluxes can change with reseating of fiber in cal source, although still couldn' | ||
| + | |||
| + | ====April 2026==== | ||
| + | |||
| + | A new iodine cell was installed. | ||
| + | |||
| + | We implemented a shutter in the direct calibration channel. | ||
| + | |||
| + | We got the repaired liquid cooled CCD back from QHY and installed it, along with the TCube Edge chiller. | ||
| + | |||
| + | When revisiting spectrograph focus, the Esatto focuser would intermittently time out. As a result, it's uncertain whether the reported focus position is repeatable, e.g., after homing the focuser. Given potential repeatability issues, the focuser was just moved until acceptable images were reached, which is at a currently reported value of 398000. After settling on that, the Esattor focuser was powered off through the Wanderer PowerBox to remove the light source. A final image at this focus is UT260424/ | ||
| + | |||
| + | {{: | ||
| + | {{: | ||
| + | |||
| + | |||
| + | |||
| ====November 2025==== | ====November 2025==== | ||
| Enclosure box installed 11/18-29. Required taking ferrule out to run through box, so location of traces, etc. changed. Installed 4 temperature sensors inside the box. Cleaned up room and put spare parts on shelf and in drawers. | Enclosure box installed 11/18-29. Required taking ferrule out to run through box, so location of traces, etc. changed. Installed 4 temperature sensors inside the box. Cleaned up room and put spare parts on shelf and in drawers. | ||
| - | Spectrograph refocussed. | + | Spectrograph refocussed. |
| + | |||
| + | Images from 433000 (left) to 438000 (right) in steps of 1000 near center of chip: | ||
| + | |||
| + | {{: | ||
| + | |||
| + | Extracted FWHM/ | ||
| + | |||
| + | |{{: | ||
| + | |||
| + | Not overly satisfying, but adopted 437000. | ||
| ====Guiding tests==== | ====Guiding tests==== | ||