on_sky_testing
Differences
This shows you the differences between two versions of the page.
| Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
| on_sky_testing [2025/11/04 23:46] – holtz | on_sky_testing [2025/11/21 15:46] (current) – holtz | ||
|---|---|---|---|
| Line 1: | Line 1: | ||
| + | ====November 2025==== | ||
| + | |||
| + | Enclosure box installed 11/18-29. Required taking ferrule out to run through box, so location of traces, etc. changed. Installed 4 temperature sensors inside the box. Cleaned up room and put spare parts on shelf and in drawers. | ||
| + | |||
| + | Spectrograph refocussed. | ||
| + | |||
| ====Guiding tests==== | ====Guiding tests==== | ||
| Line 8: | Line 14: | ||
| ====Throughput==== | ====Throughput==== | ||
| - | Spectrograph throughput at 5560-5570 | + | |
| + | ===Spectrograph throughput at 5560-5570== | ||
| + | |||
| + | For each science spectrum, the flux is extracted in each order using a boxcar extraction, and a wavelength solution from a ThAr frame is attached. We then determine the maximum flux between 5560 and 5570 Angstrom. We convert this to electrons per second for a star with m=0 using the exposure time, V magnitude of the star, and gain; we'll loosely call this throughput. This is plotted in the top panel of the following plot (other panels are not relevant here, but are the observed flux in the exposure and predicted S/N based on this): | ||
| {{: | {{: | ||
| + | Clearly, there is some variation in the throughput, which is expected because of variations in transparency and also variations in seeing, leading to different fiber losses at the focal plane. However, the exposures with iodine (+ symbols) are clearly significantly lower than those without iodine (squares) by a factor of roughly 2-3 (we've been using 2.5x the exposure time for iodine, and you can see in the middle panel that this roughly brings observed fluxes to be the same); i.e., iodine throughput is ~40% of the non-iodine throughput. | ||
| - | Stellar profiles with and without iodine | + | The horizontal lines represent throughput achieved at the Tenerife SONG node for several different stars (median of many observations), |
| + | |||
| + | ===Stellar profiles with and without iodine=== | ||
| + | |||
| + | To understand the source, we observed a star on the guide camera (off of the fiber hole) with and without iodine cell, and plot the cumulative counts in successive circular apertures. Top plot shows the two cumulative flux curves, and bottom plot the ratio. | ||
| {{: | {{: | ||
| {{: | {{: | ||
| - | Cal channel | + | The profile with the iodine |
| - | {{: | + | The pixel scale of the guide camera is about 0.16 arcsec/ |
| + | Paul also mentioned that the ends of the cell are likely to be bowed a bit, but we don't know how much image degradation this might lead to. | ||
| - | Focus run with iodine | + | ===Image quality |
| - | {{:test:iodine.png? | + | The following images shows a focus sweep of a star with the iodine |
| - | Coarse | + | {{: |
| + | {{: | ||
| + | |||
| + | The following images show coarse | ||
| {{: | {{: | ||
| {{: | {{: | ||
| + | |||
| + | ===Cal channel iodine/ | ||
| + | |||
| + | To look at the overall throughput of the iodine cell in a different way, we looked at flat fields taken with and without the iodine cell. The ratio of these is shown in the following plot. | ||
| + | |||
| + | {{: | ||
| + | |||
| + | This suggests an overall throughput of 75% or less, under the assumption that the same amount of calibration light is incident on the fiber; our calibration spot is 200 microns in size for a 50 micron hole, and we have adjusted the iodine cell so that the hole is in almost the same location when the cell is in place, but it is not exact, although we think it is close enough that roughly the same amount of flux should be going down the hole. | ||
| + | |||
| + | Interestingly, | ||