User Tools

Site Tools


song_spectrograph_assembly

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
song_spectrograph_assembly [2025/05/28 19:20] – [SONG spectrograph assembly] 172.24.37.106song_spectrograph_assembly [2025/05/30 17:55] (current) 172.59.0.80
Line 1: Line 1:
 +====== Spectrograph assembly guide ======
 ===== Room requirements/setup ===== ===== Room requirements/setup =====
  
Line 67: Line 68:
       - rotate far diagonal to direct towards diagonal under collimator for all positions of far diagonal       - rotate far diagonal to direct towards diagonal under collimator for all positions of far diagonal
       - adjust diagonal under collimator to get beam in collimator.       - adjust diagonal under collimator to get beam in collimator.
-{{ :spec:collimator.jpg?direct=300 }}+{{ :spec:collimator.jpg?direct=400 }}
  
   * With aligned collimator install flat (grating flat, FM2, and FM3 on table in desired location. Also install prism mount for convenience before adding upper table. For each of the three flats:   * With aligned collimator install flat (grating flat, FM2, and FM3 on table in desired location. Also install prism mount for convenience before adding upper table. For each of the three flats:
Line 73: Line 74:
     * Focus autocollimator to get retroreflection reticle     * Focus autocollimator to get retroreflection reticle
     * adjust tip-tilt of flat to get alignment     * adjust tip-tilt of flat to get alignment
-{{ :spec:flats.jpg?direct&300 }}+{{ :spec:flats.jpg?direct&400 }}
  
   *  set up upper table on main optical bench in correct location. Install:   *  set up upper table on main optical bench in correct location. Install:
Line 85: Line 86:
       * laser should be in center of white beam on table; if not, something is not right!       * laser should be in center of white beam on table; if not, something is not right!
       * put flat mirror on table at beam, look for reflected laser at laser source, adjust diagonal down-pointing mirror until laser reflection returns exactly to laser.       * put flat mirror on table at beam, look for reflected laser at laser source, adjust diagonal down-pointing mirror until laser reflection returns exactly to laser.
-{{ :spec:upper.jpg?direct=300 }}+{{ :spec:upper.jpg?direct=400 }} 
   * Install slit to ensure focal plane is at desired height above optical table   * Install slit to ensure focal plane is at desired height above optical table
     * measure distance from bottom of slit holder to slit (10mm)     * measure distance from bottom of slit holder to slit (10mm)
Line 123: Line 125:
 Photos of the fully assembled spectrograph at APO on May 7, 2025. Photos of the fully assembled spectrograph at APO on May 7, 2025.
  
-{{:spec:spectrograph_may07_above.jpg?400|}} +{{:spec:spectrograph_may07_above.jpg?300|}} 
-{{:spec:spectrograph_may07_low.jpg?400|}}+{{:spec:spectrograph_may07_low.jpg?300|}}
  
  
  
song_spectrograph_assembly.1748460041.txt.gz · Last modified: 2025/05/28 19:20 by 172.24.37.106