Line 33: Line 33:
 Given the fibers in the focal plane, it's not possible to illuminate all of the pixels in the detector with light coming through the fibers. One doesn't want to flat field with the calibration light because of all of the pixels with low/zero light levels. Given the fibers in the focal plane, it's not possible to illuminate all of the pixels in the detector with light coming through the fibers. One doesn't want to flat field with the calibration light because of all of the pixels with low/zero light levels.
  
-We obtained some pixel flats by removing the fiber input assembly and putting a diffuse light source (an LED lamp with a piece of paper over it) to get illumination across the whole chip:+We obtained some pixel flats by removing the fiber input assembly and putting a diffuse light source (an LED lamp with a piece of paper over it) to get illumination across the whole chip. Subsequently we achieved these using a penlight wrapped in bubble wrap which can be inserted in the upper assembly without  moving the fiber input assembly at all.
  
 {{:test:pixflat.png?direct&200}} {{:test:pixflat.png?direct&200}}