3.5-m Telescope and Instruments Highlights, 11/05/2024 – 12/02/2024
1) Overview
November has seen wintery weather with two snow storms, each with enough snow to impact operations. During the month of November we had one astronomer group visit, one visiting instrument team on site, and no class visits.
The October water leak was repaired successfully and hasn’t shown any issues since the repair. The phone line that was damaged was eventually repaired and now all the services affected by the leak are back to normal operational state.
2) Operations
3.5m Telescope: Telescope is moving as expected.
0.5m Telescope: Telescope is working well.
KOSMOS: nothing to report.
ARCTIC: had an emergency vacuum service performed due to the vacuum softening. Instrument was returned to service without any missed observing time. Instrument seems to be experiencing an increased rate of the one quadrant dropping. Investigations to determine the cause of this trend are ongoing.
Agile: Once we received the repaired camera back on site, in-lab testing revealed an issue that necessitated that we return the camera to the repair vendor to investigate the hardware that was changed. The camera control software was not able to configure the camera as expected and was reporting issues with the gain parameter setting.
ARCES: Instrument is performing nominally. IOL measurements have held steady and are still very good. Recently, the quartz flat lamp bulb housing and/or wiring was worked on to due to the lamp not turning on. A temporary repair is in place while parts are on order. Instrument flats are working with the temporary fix.
DIS: Instrument is currently up and usable. Vacuum servicing is not planned for Q4.
NICFPS: Instrument is performing nominally.
TripleSpec: Instrument is performing nominally.
APOLLO: system suffered a chiller failure due to leaks. The leaks were repaired and the system has successfully ranged to the moon after the repair. The chiller that had the issue is a “spare” chiller being used because the in-service chiller had failed. A replacement chiller is on order and will be integrated into the system when it arrives at APO.
xxx
We have some unassigned time in Q4 (Nov 23B and 28B (short), and a fair bit in December). Users should look at the schedule and follow the usual channels for requesting this time (i.e. email Russet, Amanda, Ben, Nancy and their institutional scheduler when submitting requests, and provide a proposal cover page if you don't already have a program scheduled for the current quarter).
xxx
The Q4 ARCSAT scheduled is posted. There is time available in the last week of December (but with no requests for filter changes).
xxx
xxx
xxx
xxx
Open action items from previous meetings:
Open action items from this meeting:
xxxxxx
None.
The next meeting will be on January 7, 2025 at 10:30 MST. (??)